Ivanova, Violeta and Dörnyei, Ágnes and Stefova, Marina and Stafilov, Trajče and Vojnoski, Borimir and Kilár, Ferenc and Márk, László (2011) Rapid MALDI-TOF-MS Detection of Anthocyanins in Wine and Grape Using Different Matrices. Food Analytical Methods, 4. pp. 108-115.
Preview |
Text
Ivanova-Food Anal Methods(2011) 4- 108-115.pdf Download (229kB) | Preview |
Abstract
Matrix-assisted laser desorption/ionization timeof-
flight mass spectrometry (MALDI-TOF-MS) as a new
valuable technique has been used for qualitative screening
of anthocyanins in wine and grape samples from Vranec
and Merlot variety. For that purpose, different MALDI
matrices have been tested, such as α-cyano-4-hydroxycinnamic
acid, sinapic acid, 2,5-dihydroxybenzoic acid (2,5-
DHB), and C70 fullerene. It was observed that 2,5-DHB
was superior with respect to all the matrices used in the
study. Furthermore, fullerene was, for the first time, tested and used as a possible matrix for MALDI-TOF-MS
identification of anthocyanins in wine and grape samples.
The results obtained showed identification of aglycone
forms of anthocyanins only, since the higher laser energy
applied for ionization of the fullerene molecules led to great deal of fragmentation of the sugar moiety from glucosides, acetylglucoside, and p-coumaroylglucoside groups from the corresponding anthocyanin molecules. Additionally, liquid chromatography/electrospray ionization mass spectrometry has also been applied for anthocyanin characterization.
Item Type: | Article |
---|---|
Subjects: | Agricultural Sciences > Agricultural biotechnology Natural sciences > Chemical sciences |
Divisions: | Faculty of Agriculture |
Depositing User: | Violeta Ivanova Petropulos |
Date Deposited: | 04 Nov 2012 17:50 |
Last Modified: | 28 Oct 2013 12:39 |
URI: | https://eprints.ugd.edu.mk/id/eprint/305 |
Actions (login required)