Spirevska, Ilinka and Soptrajanova, Lidija and Gulaboski, Rubin (2000) Square-Wave Voltammetric Method for Determination of Aconitic Acid. Analytical Letters, 33 (5). pp. 919-928. ISSN 0003-2719
Preview |
Text
__ugd.edu.mk_private_UserFiles_rubin.gulaboski_Desktop_GULABOSKI-MY pdf PUBLICATIONS_Fumarna-Anal Letters 2001.pdf Download (217kB) | Preview |
Abstract
A square-wave voltammetric method at a hanging mercury drop electrode for quantitative determination of aconitic acid in aqueous solution is developed. Aconitic acid is an electroactive compound in acid and neutral medium. Its voltammetric response depends on the type as well as the pH of the supporting electrolyte. The optimization of experimental parameters for quantitative determination of aconitic acid was achieved. The quantitative determination of aconitic acid was carried out in o. 1 mol/L HCl as a supporting electrolyte. The effect of interferences is studied. The detection limit of aconitic acid is 0.184 μg/mL.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | square-wave voltammetry, biosensors, aconitic acid, electrochemistry |
Subjects: | Agricultural Sciences > Agricultural biotechnology Natural sciences > Chemical sciences Natural sciences > Physical sciences |
Divisions: | Faculty of Agriculture |
Depositing User: | Rubin Gulaboski |
Date Deposited: | 31 Oct 2012 19:23 |
Last Modified: | 31 Oct 2012 19:23 |